Warnings; Error Reports - B.Braun Aesculap GN200 Instrucciones De Manejo

Ocultar thumbs Ver también para Aesculap GN200:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 58
7.2

Warnings

A warning is indicated as follows:
Individual tone (different from a failure, an error or normal HF opera-
tion)
Warning message in the display 5 on the front side of the HF device
Warning message
Generator Cooling
Please Wait
Generator Too Cold
Please Wait
7.3

Error reports

An error is indicated as follows:
Clear tone sequence (different from a failure or normal HF operation)
Error message in the display 5 on the front side of the HF device
Error code
Error report
E001
ERROR E001
Instrument RF-ON Button Error
Release Instrument RF-ON Button
Remove and reattach instrument
E002
ERROR E002
Instrument Error
Remove and reattach instrument
- or -
Remove and replace instrument
E003
ERROR E003
Footswitch Error
Release footswitch
Remove and reattach footswitch
Significance
Excess generator temperature
Generator must first warm up
Significance
"HF activation (RF-on/off) 2" instrument
switch pressed while the instrument is
connected to the HF device
HF device cannot find instrument
- or -
Instrument is defective and must be
replaced
Foot switch is pressed while the foot
switch is connected to the HF device
Remedy
Check the airflow around the generator
Do not switch off the generator whilst it is warming up
Remedy
Release the "HF activation (RF on/off) instru-
ment switch 2".
Remove instrument and reconnect it
Remove instrument and reconnect it
- or -
Replace instrument
Release foot switch
Remove foot switch and reconnect it
15

Publicidad

Tabla de contenido
loading

Tabla de contenido