PROPRIETARY AND CONFIDENTIAL DRAFT: octobre 11, 2012
Control unit
Front view of the control unit of the Primo
Control
panel
Characteristics
Model
Dimensions
Weight
Power supply
Apparent power
Maximum energy consumed
Operation of the device
Sleeve material
Device : sound pressure level /
sound power (ISO 3744)
Alarm : sound pressure level /
sound power (IEC60601-1-8)
Fuses
Compression
Compressor flow rate
Device classification according
to IEC 60601-1
Degree of protection provided
by the cover (IEC 60529)
Protection against inflammable
anesthetic mixtures
Battery life
170924(2) - EN - Primo
P02033
220-240V
50/60 HZ
20 VA
16.4 Wh
T160 mA
0 - 180 mbar
10 l/min
IP21 : protected against access to hazardous parts with a
finger and against vertically falling water drops.
Not for use with flammable anesthetics.
User manual
™
Description
P02047
P02034
15x29x12 cm / 6x11.5x5''
3.5 kg / 8 lb
220-240V
120V 60HZ
50/60HZ
20 VA
32 VA
16.4 Wh
16.4 Wh
Continuous
ABS PC V0
41 dB(A) / 52 dB(A)
56 dB(A) / 67 dB(A)
T160 mA
T 200mA
0 - 180 mbar
0 - 215 mbar
10 l/min
12 l/min
Class II
1 hour
™
P02044
220-240V
50/60HZ
20 VA
16.4 Wh
T160 mA
0 - 180 mbar
10 l/min
Page 7