PROPRIETARY AND CONFIDENTIAL DRAFT: octobre 11, 2012
10. Connect the power cord connector to the plug (G).
11. Turn the lock ring (H) clockwise to lock the coupling.
12. Check that the CPR valve is closed.
13. Make sure that the power cord is not damaged and connect it to the control
unit.
14. Attach the power cord to immobile parts of the bed frame using the cable
clamps. Make sure that the cord will not be jammed or drag along the floor.
Check that the clamps of the power supply cord are attached to the main frame, and
not to articulated parts, such as the head or foot sections. Failure to follow this
recommendation could result in damage to the equipment.
Take care not to damage the power cord when moving the bed. Failure to follow this
recommendation could result in corporal injury.
15. Connect the power cord to the wall outlet.
16. Press the
The yellow visual alarm signal
When connecting to the mains for the first time, the mattress automatically
inflates. This takes about 15 minutes.
Do not place the patient on the mattress while the indicator light is flashing during
the initial inflation phase.
Placing a cotton sheet on the mattress will improve patient comfort and make it
easier to administer care.
When the mattress is in use, make sure that the patient's sacrum is centered and
positioned at the I-mmersion
Page 18
button on the control unit to switch on the device.
Therapy symbol located on each side of the mattress.
comes on.
170924(2) - EN - Primo
User Manual