Manufacturer's declaration - Electromagnetic Compatibility
Electromagnetic emissions compliance
Manufacturer's recommendations and declaration – electromagnetic emissions
The Primo ™ is intended for use in the electromagnetic environment specified below. The customer or
the user of the Primo ™ should make sure it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Emissions of voltage
fluctuations/flicker
IEC 61000-3-3
Electromagnetic immunity compliance
Manufacturer's recommendations and declaration – electromagnetic immunity
The Primo ™ is intended for use in the electromagnetic environment specified below. The customer or
the user of the Primo ™ should make sure it is used in such an environment.
Test level
Immunity test
IEC 60601
Electrostatic
discharge
±6 kV on contact
(ESD)
±8 kV in the air
IEC 61000-4-2
Fast electric
±2 kV on power
transients in
supply lines
bursts
±1 kV for the
IEC 61000-4-4
input/output lines
±1 kV between line(s)
Shock waves
±2 kV between line(s)
IEC 61000-4-5
and earth
Magnetic field
at the
frequency of
the mains
3 A/m
power supply
(50/60 Hz)
IEC 61000-4-8
170924(2) - EN - Primo
Compliance
Electromagnetic environment - Recommendations
The Primo ™ only uses RF energy internally.
Consequently, it only produces very weak RF emissions
Group 1
that are unlikely to cause interference with nearby
electronic equipment.
Class B
The Primo ™ is suitable for use in all establishments,
Class A
including domestic premises and premises directly
connected to the public low-voltage power supply in
residential buildings.
Compliant
Compliance
±6 kV on contact
±8 kV in the air
±2 kV on power
supply lines
±1 kV for the
input/output lines
±1 kV between line(s)
±2 kV between line(s)
and earth
3 A/m
User Manual
™
Electromagnetic environment -
Recommendations
Floors should be made of wood,
concrete or ceramic tiles. If the floor is
covered with a synthetic material,
relative humidity must be at least
30%.
The quality of the electric power
supply should be the equivalent of
that found in commercial or hospital
environments.
The quality of the electric power
supply should be the equivalent of
that found in commercial or hospital
environments.
The magnetic field at the frequency of
the mains supply must be of a level
that is characteristic of a place
located in a commercial or hospital
environment.