Recommended Separation Distances - Hill-Rom Primo P02033 Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 312
PROPRIETARY AND CONFIDENTIAL DRAFT: octobre 11, 2012
Manufacturer's recommendations and declaration – electromagnetic immunity
The Primo ™ is intended for use in the electromagnetic environment specified below. The customer or
the user of the Primo ™ should make sure it is used in such an environment.
Test level
Immunity test
IEC 60601
• <5% U
(dips >95% of U
Voltage dips,
for 0.5 cycle
brief outages
• 40% U
and variations
(dips = 60% of U
in voltage on
for 5 cycles
the electric
• 70% U
power supply
(dips = 30% of U
lines
for 25 cycles
IEC 61000-4-
• <5% U
11
(dips >95% of U
for 5 seconds
a. U
is the nominal value of the supply voltage applied during the test.
T

Recommended separation distances

RF communication equipment (including peripherals such as antenna cables and external
antennas) should be used no closer to any part of the Primo
Hill-Rom , than the recommended separation distances shown in the following table.
Otherwise, degradation of the performance of this equipment could result.
Wireless communication system
TETRA 400 (limited to 10 W ERP*)
Public safety(460-470 MHz), GMRS 460 (limited to 5 W ERP*)
GMRS 460 (limitede to 2 W ERP*)
GSM 850, GSM 900, RFID 868 MHz, TETRA 800 (limited to 2 W
ERP*)
GSM 1900 (limited to 1 W ERP*)
WLAN 802.11a 5 GHz (limited to 1W ERP*)
iDEN 820, CDMA 850, GSM 1800, CDMA 1900 (limited to 0,6 W
ERP*)
FRS 460 (limited to 0,6 W ERP*), PMR 446
UMTS, DECT (limited to 0.25 W ERP*)
*:Effective Radiated Power
Page 40
Compliance
• <5% U
a
T
)
(dips >95% of U
T
for 0.5 cycle
• 40% U
T
)
(dips = 60% of U
T
for 5 cycles
• 70% U
T
)
(dips = 30% of U
T
for 25 cycles
• <5% U
T
)
(dips >95% of U
T
for 5 seconds
Electromagnetic environment -
Recommendations
T
The quality of the electric power
)
T
supply should be the equivalent of
that found in commercial or hospital
T
environments.
)
T
If the user of the Primo ™ requires the
appliance to be in continuous use
T
)
during power cuts, it is advisable to
T
power the Primo ™ from an
uninterruptible power source or a
T
)
battery.
T
, including cables specified by
170924(2) - EN - Primo
Separation distance
(m)
0,3
0,2
0,1
0,3
0,2
0,7
0,2
0,1
0,1
User Manual

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para Hill-Rom Primo P02033

Este manual también es adecuado para:

Primo p02034Primo p02044Primo p02047

Tabla de contenido