WARNING: Risk of explosion and fire due to use of
unsuitable liquids
Use only water for disinfection.
Maintain the following times when disinfecting:
Wattage [W]
850 – 950
600 – 850
CAUTION: Risk of burning
At the end of the disinfection process, the parts may still
be very hot. Therefore, it is essential to wait for the
cooling time to pass before taking the parts out.
Info:
Information about the amount of water required and about the disinfec-
tion process is included in the instructions for use of the devices in
question (disinfector and microwave).
Place the components in the disin-
fector with enough space between
them. If possible, place the parts in
the appliance so that the steam
created by the microwave action
can flow through the parts to the
greatest effect.
Disinfection time
4 min.
6 min.
Cooling time
2 min.
2 min.
en