Determination of the aerosol output rate
Fill the nebuliser with 2 ml of a 1% test solution. Attach the nebu-
liser to the Breath Simulator for at least 1 minute.
Aerosol output rate in [ml/min] for minimum flow (3.0 l/min) and
maximum flow (6.0 l/min):
Min. flow (3.0 l/min)
Max. flow (6.0 l/min)
(Averages from each of 2 separate measurement series with 3 nebuliser/
compressed air combinations)
We reserve the right to make technical changes.
Information as of: 2018-11
3) Quantity of aerosol that is output by the nebuliser system per unit of
time
3)
0.096 ml/min
0.144 ml/min
en
69