®
– Polytron
5700 IR with PIR 7000 is an explosion proof
instrument for the continuous monitoring of combustible
gases and vapors containing hydrocarbons.
®
– Polytron
5720 IR with PIR 7200 is an explosion proof
instrument for the continuous monitoring of carbon
dioxide.
3.5
Approvals
3.5.1 ATEX, IECEx, UL, CSA
See a sample of approval label on the fold-out page. There is
also a printout of the approval label inside the shipping box of
the instrument.
3.5.2 Marking
The marking is reproduced on a separate piece of paper
shipped with the instrument.
Serial Number key: The third letter of the serial number
specifies the manufacturing year: M = 2019, N = 2020, P =
2021, R = 2022, S = 2023, T = 2024, U = 2025, W = 2026, X =
2027, Y = 2028, Z = 2029, etc. (Letters G, I, O, Q are omitted)
Example: Serial Number ARMB-0001: the third letter is M,
which means that the unit was manufactured in 2019.
4
Explosion proof (Ex d) installation
- without Docking Station
Obey the safety statements mentioned in 1 Safety-related
information!
4.1
Mechanical installation without
Docking Station
The instrument can be mounted for measuring gases in
ambient air or inside of a duct/pipe.
1. Use M6 (1/4'') bolts with hex socket caps to mount the
enclosure to one of the following options.
Option
Mounting on a flat surface
Mounting on a pole
Mounting on a duct/pipe
4.2
Electrical installation without
docking station
For recommended tightening torques and admissible wire
specifications see 17.3 Cable specifications.
Dräger Polytron
Accessory
Drilling template: 4544299
®
Additionally for Polytron
57x0 IR: Spacer 6812617
Pole mount kit: 4544198
Duct mount kit: 6812725
®
For Polytron
57x0 IR:
6812300
®
5000 Series
Explosion proof (Ex d) installation - without Docking Station
4.2.1 Wiring tables
4-20mA / Power
Pin
Mark
Function
1
4-20 /
4 to 20 mA signal (Operation as source or
S
sink)
2
V-
3
+
V+
®
Polytron
5100 EC 4-20mA / Power
Refer to figure H on the fold-out page.
Pin
Mark
Wire
color
1
4-20 /
brown 4 to 20 mA signal (Operation as
S
2
black
3
+
red
Relay connector
The relay labels (NO, COM, NC) represent the default state
(normally energized) of all relays while the instrument is
powered.
Pin
Mark
1
NO
FLT
2
COM
3
NC
4
NO
A2
5
COM
6
NC
7
NO
A1
8
COM
9
NC
4.2.2 Figures on the fold-out page
Fig-
Connecting diagram for instruments
ure
G
Enclosure grounding
G1
PCB unit Polytron
G2
PCB unit Polytron
H
Enclosure grounding and internal wiring of Polytron
1)
5100 EC
I1
4-20 mA (Current source)
I2
4-20 mA (Current sink)
I3
Stand-alone, relay only
H1
2-wire installation of Polytron
J
Relay board installation
Function
source or sink)
V-
V+
Relay
Fault
Normally Open
Fault
Common
Fault
Normally Closed
A2
Normally Open
A2
Common
A2
Normally Closed
A1
Normally Open
A1
Common
A1
Normally Closed
1)
®
57x0 IR
®
®
5200 CAT / Polytron
5310 IR
®
5100 EC
en
®