At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may
EN
need to perform preventive maintenance checks more frequently based on your level of product usage.
Remove product from service before performing preventive maintenance. Preventive maintenance should only be
performed by trained or certified personnel.
Note: Clean and disinfect the exterior of the mattress before inspection, if applicable.
Inspect the following items:
All welds and all fasteners are secure
Tubing or sheet metal for bends or breaks
Casters are free of debris
Casters are secure and swivel properly
Casters lock securely by depressing the brake pedal
Steer caster latches properly
Siderails move and latch properly
Fowler operates properly
Fowler (back rest) slide operates properly
Litter Up/Down operates properly
Foot section operates properly
Footrests operate properly
Trend operates properly
IV pole is intact and operating properly
Mattress cover after each use
Mattress cover for rips or cracks
Headboard and siderail panels for cracks or splits
All covers are not damaged and are not producing sharp edges
Night light operating properly
CPR release operating properly
All siderail motion functionality
Nurse call functionality (Optional equipment)
9V Nurse call battery (replace annually) (Optional equipment)
Lubricate where required
Pendant for any physical damage
Power cord not frayed
Cables not worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 100 mΩ (milliohms)
Current leakage not more than 300 µA (microamps)
Ground chains are clean, intact, and have at least two links touching the floor
Enclosure is free from wear, tear, stresses and mechanical damage
No rust or corrosion of parts
Labels for legibility, proper adherence, and integrity
Product Serial Number:
Completed by: ______________________________________ Date:_____________
1-42
Preventive maintenance
4701-209-0 01 REV A