3. Clean
• Prepare cleaning solution using alkaline detergent or neutral pH
enzymatic detergent per manufacturer's directions for correct
temperature, water quality (i.e. pH, hardness) and product
concentration/dilution.
• Fully immerse attachment in cleaning solution in a suitable container
and manually agitate for 2 minutes.
• Use a sponge, soft lint-free cloth or soft bristle brush to remove all
visible soil.
• Gently use a brush to clean cannulations of attachment while
immersed in cleaning solution. Use ACB for all attachments except
MA-D20-G1 and Minimal Access Attachment Bearing Sleeves. Use
SDCB for MA-D20-G1 and Minimal Access Attachment Bearing
Sleeves. If attachment has a release sleeve (Figure 1), move sleeve
and brush exposed area.
Note: Brushes may be inserted from distal or proximal end of
attachments and bearing sleeves.
Warning: ACB and SDCB are single use and should only be used to clean
one set of attachments.
Caution: For MA-D20-G1 and Minimal Access Attachment Bearing
Sleeves use SDCB only. Do not force ACB or SDCB into or through
attachments or bearing sleeves. Do not use pipe cleaners.
Figure 1
4. Rinse Thoroughly
• Rinse device thoroughly using running hot deionized, distilled, or
purified water for a minimum of 2 minutes.
• Use a syringe or pipette filled with hot deionized, distilled or purified
water to flush lumens or channels.
• Manipulate any moving parts such as release sleeves (Figure 1).
Caution: Do not use high pressure water or air. Do not rinse attachment
with saline solution.