Option 1: 05.001.078, Lubricant for Anspach Systems, 110 ml
Warning: Do not use Gravity Air Displacement Steam Autoclave
Sterilization Process.
• For the following attachments apply one (1) spray pump
spray to the proximal end of the attachment.
SHORT-G1
SHORT-HD-G1
MEDIUM-G1
MEDIUM-HD-G1
• For the following attachments apply one (1) spray pump
spray to the distal end of the attachment and one (1) to the
proximal end of the attachment.
MA-D20-G1
CSR60-G1
• Applying excess amounts of lubricant will cause the lubricant
to drip from the attachment. Clean off any excess lubricant.
Option 2: Non-silicone based medical lubricant
• Prepare a lubricating solution of instrument milk (non-silicone
based medical lubricant) per the manufacturer's directions.
• Fully immerse attachment in lubricating solution, at room
temperature, in a suitable container and agitate for
15 seconds.
• Remove attachment and allow it to drain completely until no
visible droplets are coming from it.
Caution: Do not rinse out instrument milk (lubricant). Do not apply
mineral oil or other lubricants, which may cause attachment to
overheat.
7. Go to Sterilization section for further instructions.
Consoles
Caution: Do not immerse or sterilize. Do not allow liquid to enter
console.
Note: Consoles are non-sterile capital equipment and shall always
be maintained outside the Operating Room Sterile Field.
LONG-G1
LONG-HD-G1
MIA16-G1
CRANI-A-G1
QD8-G1
QD8-S-G1
QD11-G1
QD11-S-G1
CRANI-L-G1
CRANI-P-G1
MA-10S
MA-10C
QD14-G1
QD14-S-G1
MA-15S
MA-15C
MA-15ST
MA-19ST
5