MA-D20-G1
CSR60-G1
• Applying excess amounts of lubricant will cause the lubricant
to drip from the attachment. Clean off any excess lubricant.
Option 2: Non-silicone based medical lubricant
• Prepare a lubricating solution of instrument milk (non-silicone
based medical lubricant) per the manufacturer's directions.
• Fully immerse attachment in lubricating solution, at room
temperature, in a suitable container and agitate for 15
seconds.
• Remove attachment and allow it to drain completely until no
visible droplets are coming from it.
• Return attachments to basket for further processing.
Caution: Do not rinse out instrument milk (lubricant). Do not apply
mineral oil or other lubricants, which may cause attachment to
overheat.
5. Go to Sterilization section for further instructions.
Note:
• The clinical processing instructions provided have been
validated by DePuy Synthes Power Tools for preparing a
non-sterile medical device; this instruction is provided in
accordance with ISO 17664 and ANSI/AAMI ST81.
Sterilization
Before sterilizing:
• Ensure sterilization equipment is in proper working order as
specified by manufacturer.
• Assure equipment manufacturer's instructions are properly
employed by trained and qualified personnel.
• Assure actual cycle employed has been properly validated
for the device(s)/load configuration being processed and
appropriate sterilization indicator devices are included for
each process and cycle.
QD8-G1
QD11-G1
QD8-S-G1
QD11-S-G1
QD14-G1
QD14-S-G1
13