Section 6 Electrosurgical Precautions
The safety and effectiveness of electrosurgery is dependent to a large degree upon the skill of the user/
operator . It is important that the user/operator read, understand, and follow these operating instructions
supplied with the CooperSurgical LEEP PRECISION Generator as well as thoroughly understand the
principles and use of electrosurgical systems .
Electrosurgery uses radio-frequency energy to cut and coagulate tissue . Because of the sparking and heat
associated with electrosurgery, do not use with flammable anesthetics or other flammable gases, near
flammable fluids or objects, or with oxidizing agents .
Skin-to-skin contact, for instance between the patient's arm and body, should be avoided by the
placement of an appropriate separating device, such as two to three inches of dry gauze . This will reduce
the potential for alternate site burns .
If monitoring, stimulation, imaging, or similar devices are used simultaneously with electrosurgery, the
monitoring electrodes must be placed as far as possible from the electrosurgery site and the Patient
Return Electrode . Position the Patient Return Electrode close to the electrosurgery site, for example, on
the thigh when treating the cervix . NOTE: Monitoring Needle Electrodes are not recommended .
WARNING