Section 17 EMC Compliance Information for the LEEP PRECISION Generator
• Medical electrical equipment use requires special precautions regarding EMC and must to be installed and put
into service according to the EMC information provided in the Accompanying Documents .
• Portable and mobile RF communications equipment can affect medical electrical equipment .
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC EMISSIONS
The CooperSurgical LEEP PRECISION Generator is intended for use in the electromagnetic environment
specified below . The customer or the end user of the CooperSurgical LEEP PRECISION Generator should assure
that it is used in such an environment .
Emissions Test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
Compliance
Electromagnetic Environment Guidance
CooperSurgical LEEP PRECISION Integrated
Systems use RF energy only for their internal
Group 1
function . Therefore, RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment .
Class A
CooperSurgical LEEP PRECISION Integrated
Systems are suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
Class A
domestic purposes .
Complies
26