Specifications; Environment Conditions - Cooper Surgical LEEP PRECISION LP-20-120 Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 59
Section 16 Specifications
Size (W x D x H): . . . . . . . . . . . . . . . . . . . . . . . . . . .9 .38 inches x 11 .63 inches x 4 .25 inches (23 .8 cm x 29 .5 cm x 10 .8 cm)
Weight: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 pounds, 9 ounces (5 .25 kg)
Electrical:
Input Voltage
Maximum Current
VA Ratings
Power Line Leakage
Fuses (5 x 20 mm)
High Frequency Output*:
Rated output power is 100 watts, 495 kHz ± 5% into 500 ohms, monopolar only .
Minimum & Maximum
Mode
Power Settings
CUT
10 to 100 watts
BLEND
10 to 100 watts
COAG
10 to 100 watts
*Accuracy of output control setting:
For output powers in excess of 10 watts, the actual power as a function of the load resistance and output control
setting shall not deviate from that shown in the diagrams shown on pages 23 and 24 by more than ± 20 percent .
RF Leakage: 100 MA MAXIMUM AT MAXIMUM POWER SETTINGS
WEEE Directive:
In order to preserve, protect and improve the quality of the environment, protect human health and
utilize natural resources prudently and rationally – do not dispose of waste electrical or electronic
equipment (WEEE) as unsorted municipal waste . Contact local WEEE disposal sites .
Classification:
Model
LEEP PRECISION Generator
• Do not get fluid into the LEEP PRECISION Generator . Should any liquid or solid object fall into the unit, unplug
the unit and call Technical Support .
• The LEEP PRECISION Generator is suitable for intermittent operation with a two-minute ON and six-minute
OFF duty cycle .
• The LEEP PRECISION Generator is classified as normal equipment (IPX0) according to protection against
ingress of water .

Environment Conditions:

Environmental Temperature
Relative Humidity
Air Pressure
120 VAC
3 .15 amps
132 VA
Less than 50 micro-amps
Two T 3 .15A 250V, Slow-Blow,
(Littelfuse
02183 .15 or Equivalent)
®
Crest
Maximum Open
Factor
Circuit Voltage
1 .2
2 .5
6 .5
Safety Class
I
Use
between +10 °C and +40 °C
between 10% and 90%
between 700 hPa and 1060 hPa
230 VAC
1 .55 amps
242 VA
Less than 50 micro-amps
Two TT 2 .5 A 250 V, Super Slow-Blow,
(Bussmann
600
600
4000
Type
BF
Shipping and Storage
between +10 °C and +40 °C
between 10% and 90%
between 700 hPa and 1060 hPa
23
GMD-2 .5 A or Equivalent)
®

Publicidad