KPS-PA50 MINI • Digital clamp meter
• Press the "Hz/%" button again, meter will be back to the condition of the
voltage or current measuring.
NOTE:
During working at maximum or minimum value measuring function, the meter
can't be changed into frequency or duty cycle measuring mode.
4.4 Switching maximum or minimum value
• At all ranges, press the "MAX/MIN" button one time, the meter can be set
to maximum value measuring mode; press the button twice, the meter can
be set to minimum value measuring mode; press the button three times, the
meter will get back to normal test mode, and the maximum and minimum
value will be recorded by the chip.
• Press the "MAX/MIN" button more than one second or press the "RAN" but-
ton, the meter will get back to normal test.
Note:
During measuring maximum or minimum value, the meter will be set to manual
mode automatically.
During working at frequency or duty measuring function, the meter can't be
changed into maximum or minimum value measuring mode.
4.5 Switching functions
• AC A range
Press the "SEL" button, the meter will get into inrush current test mode.
If press the "SEL" button again, the meter will get into inrush current test mode
again.
Press the "SEL" button more than one second or press the "RAN" button, the
meter will get back to normal test mode.
• DC A range
Press the "SEL" button, the meter will get into zero.
If press the "SEL" button again, the meter will get into zero again.
Press the "SEL" button more than one second or press the "RAN" button, the
meter will get back to the normal test mode.
Press the "SEL" button to switch between AC and DC measurement at the
voltage ranges.
Press the "SEL" button to switch among resistance, diode and continuity rang-
es.
39