Location; Electrical Installation - Selecta STERIL- Food 20L Manual De Instrucciones

Autoclave para esterilización de conservas y liquidos en botes cerrados
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES CODIGO 80282 REV D
Left 50 cm free around autoclave:
Left a space of 50cm
J
.P. SELECTA s.a.u.
DIC/2018 (Sujetas a modificaciones sin previo aviso)

4 Location

A tap water and waste outlet will be need to install the STERIL-FOOD
autoclave. This draw shows the suggested placement and level that
and convenient.
Front

Electrical installation

Check the site electrical installation where the autoclave will be installed.
For your safety, we recommend to supply the autoclave with a magnetothermic
and a differential.
Autoclave
size
20L
Water inlet installation
In areas with "hard" water (conductivity above 100uS / cm) it is necessary to
use a water demineralization equipment.
The water inlet must have a pressure between 3 and 4 bar.
If it is necessary, you must install:
-
A pressure group, if the pressure of the water network is less than 1 bar.
-
Hand-operated switch if the water network pressure is higher than 4 bar.
Installation of a drain
Located at ground level or below.
The material must resist temperatures of 100ºC.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
AUTOCLAVE
Water inlet
Purge steam outlet
Steam & water outlet
Waste: Undergroud or ground level.
Power
Potency
(W)
230VAC
2800
Pag.: 30
Water tab
(Above autoclave
water inlet)
2bar minimum
pressure
WASTE
Magnetothermal
16A
Tel 34 937 700 877 Fax 34 937 702 362