Factory Set Parameters - Selecta STERIL- Food 20L Manual De Instrucciones

Autoclave para esterilización de conservas y liquidos en botes cerrados
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80282 REV D

15 Factory set parameters

Parameter
A
=
B
=
C
=
D
=
E
=
F
=
G
=
H
=
I
=
J
=
K
=
L
=
M
=
N
=
O
=
P
=
Q
=
R
=
S
=
T
=
U
=
V
=
W
=
X
=
Y
=
Z
=
J
.P. SELECTA s.a.u.
Default value
120
s
10
s
10
s
5
s
0
s
-0.6
bar
0.3
min
-0.6
bar
4
min
+0.50
bar
2
s
0.10
bar
120
min
+0.05
bar
-0.70
bar
60
min
01
min
-0.20
bar
-0.20
bar
040
ºC
085
ºC
002
ºC
102
ºC
081
s
020
ºC
+0.20
bar
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
DEC/2018
(It can be modified without notice)
Definition
Maximum water fill time before raising the alarm.
Purge cycle time.
Purge time in minutes.
Purge cycle, Closed valve time (s).
Purge initial temperature.
Page: 51
Tel 34 937 700 877 Fax 34 937 702 362

Publicidad

Tabla de contenido
loading

Tabla de contenido