INSTRUCTION MANUAL CODE 80282 REV D
Parameter
Default value
A (s)
120
B (s)
10
C (s)
3
G (minute)
1
K (s)
2
M (minute)
120
N (relative bar)
0.05
S (relative bar)
-0.20
U (ºC)
85
V (ºC)
2
W (ºC)
98
X (s)
30
Y (ºC)
20
J
.P. SELECTA s.a.u.
DEC/2018
9.7.2 Parameters involved in the canned autoclave ope-
ration
There parameters should only be changed following instructions from Technical
Service.
Min
Max
Definition
0
125
Maximum water fill time before raising the alarm
0
70
Water filling extra time at the beginning (after lever sensor activation)
10
99
Purge cycle
1
60
Purge time
1
9
Purge valve closed time
1
120
Maximum time to reach setpoint before giving the alarm
0
0.40
Maximum pressure to allow door opening
-0.50
00
Minimum safety pressure to end of cycle
40
100
Maximum temperature to allow door opening
0
10
Offset liquids probe temperature
90
115
Start purging temperature
80
100
Atmospheric purge time (seconds x3)
20
90
Preheat temperature water
9.7.3 Optimal use of autoclave
Parameters may in some occasions optimize the autoclave performance.
«Purge» parameters (removing the air inside the autoclave).
• For pasteurization cycles: (temperature below 100ºC) the «purge» is of
little importance to reach pasteurization temperature. So it can be reduced to 1
minute and 90ºC.
C= 10s ;
• For medium temperatures: up to 115ºC a longer air removal phase is
needed.
C= 10s
;
For high temperatures: up to 121ºC a longer air removal phase is needed.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
G=1 minute
;
G =1 minute
;
Page: 43
K= 2s
;
W = 90 ºC
K=2s ;
W = 101 ºC
Tel 34 937 700 877 Fax 34 937 702 362