INSTRUCTION MANUAL CODE 80282 REV D
ERROR 23
POWER FAIL
ERROR 30
LEVEL ERROR
ERROR 31
PURGE FAIL
ERROR 32
DESPRESS FAIL
ERROR 33
EXT. PROBE FAIL
ERROR 34
INTERN. UNDERTEMP
«C» Led state
«Eject» key
"Name" cannot be empty
FAT32
Screen "Properties" of USB pen drive
J
.P. SELECTA s.a.u.
DEC/2018
It indicates the interruption of electric fluid during the cycle execution.
Check logs to consider re-sterilization samples.
Water level detection failure.
Check the levels operation.
Failure during PURGE phase.
Failure on pressure measurement.
Heart temperature (control sample) is out of range +5, -2ºC with respect to the
autoclave probe (tank bottom) for more than 5 minutes.
The autoclave probe reading (tank bottom) is lower than 8ºC.
11 USB datalogger
The «USB DATA LINK» is a data acquisition device (datalogger) that generates a
report from the sterilization cycle and stores the readings on a USB storage device.
Logger mounted in the autoclave is product code:
Device formatting type (USB pen drive):
Operation
Installation:
• Follow the instructions in the autoclave to activate the printer and the sending
interval of the readings.
Operation:
• Connect the «USB pen drive» to the USB Data Link. Led «C» blinks and stays
fixed light. This indicates that there is a proper recognition of the USB pen drive.
Start the autoclave, and when sending the reading to the USB Data Link, led
«C» makes a quick blinking.
At the end of the autoclave cycle, press «Eject» to extract the «USB pen drive».
Connect the «USB pen drive» to a computer to download the data file.
Important note: The «USB pen drive» must have a device name. It cannot
be blank.
The file created is text type, that can be opened with WordPad or WORD.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
4120132
FAT 32
Tel 34 937 700 877 Fax 34 937 702 362
Page: 45