High Priority Alarms - GCE Zen-O lite RS-00600 Manual Del Usario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 176
9 . 1 . 1 . Critical High Priority Alarms
IMPORTANT: These alarms will disable the device immediately.
ALARM
MESSAGE
Battery needs charging.
Battery is not an approved battery.
Service required.
*Value: 01-20

9 . 1 .2 . High Priority Alarms

IMPORTANT: These alarms will allow the device to continue operating.
ALARM
MESSAGE
Device is unable to maintain oxy-
gen purity.
Estimated battery life less than 17
minutes.
Service required.
*Value: 21-50
DESCRIPTION
DESCRIPTION
25/340
ACTION
Recharge the battery pack by plug-
ging in to the power supply. Ensure
all connections are made securely.
Replace battery with an approved
battery.
Contact your distributor.
ACTION
Be sure air inlet/outlet has not
been blocked. If alarm persists,
contact your distributor.
Charge the battery pack by plug-
ging in to power supply.
Important: The message will be au-
tomatically cleared when plugged
in to power supply.
Contact your distributor.
EN

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para GCE Zen-O lite RS-00600

Este manual también es adecuado para:

Zen-o lite rs 00600c

Tabla de contenido