The CapsoCam Plus is intended for use in the electromagnetic environment specified below.
The customer or the user of the CapsoCam Plus should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
12
Guidance and manufacturer's declaration – electromagnetic immunity
IEC 60601
Test Level
3 Vrms
150 kHz to 80 MHz
10 V/m
80 MHz to 2.7 GHz
Document Number: IFU-2320 Rev: J Revision Date: 08/2020
(Table 3)
Compliance level
Not Applicable
Portable and mobile RF communications
equipment should be used no closer to
any part of the EQUIPMENT, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
20 V/m
Recommended separation distance
d = 0.18√P 80 MHz to 800 MHz
d = 0.35√P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey
compliance level in each frequency
range.
Electromagnetic environment –
guidance
, should be less than the
a
b