Select by means of the "Central control" [8] the program
number under which the settings must be stored. The display
shows the already "old" stored settings under that number.
•
Operate the "Accept key" [7]
•
In the display appears "Overwrite ?" Select by means of the
"Central control" [8] "Yes" if you wish to continue with the
procedure or "No" if you wish to cancel the procedure.
•
Operate the "Accept key" [7] once again. Two short beeps
indicates that the treatment suggestions are stored into the
memory.
Storage into the protocol memory
•
Set the required therapy mode.
•
Set the treatment time.
After adjustment of the treatment parameters the bottom line of the
display can be filled. This line is used for additional, advise,
information regarding the treatment. This information is also stored.
•
Indicate which radiator type must be used.
•
Indicate which dosage must be used (see §6.2. item 14). It
is possible, if necessary to set two values.
•
Indicate in which range (in Watt) the power setting will
probably be set (minimum and maximum value).
•
Operate "Program store key" [2]. In the upper right corner
ofthe display [12] the program number flashes (e.g. #1).
•
Select by means of the "Central control" [8] in which cluster (A
till E) the treatment suggestion must be stored.
•
Move the cursor one position to the right by means of the
"Display parameter selectors" [4].
•
Select by means of the "Central control" [8] the program number
under which the settings must be stored. The display shows the
already "old" stored settings under that number.
•
Operate the "Accept key" [7]