Classification
International safety standard IEC 60601-1
Non ionising radiation
Safety class 1
The unit is equipped with a safety earth and must be connected to a mains plug with earth contact.
Type B
Technical modifications reserved.

11.3 EMC Details

Medical electrical devices such as the RADARMED are subject to special precautions with regard to
electromagnetic compatibility (EMC) and must be installed and commissioned in accordance with the
EMC advice given in the instructions for use and accompanying documents.
Portable and mobile RF communication systems (e.g. mobile phones) may interfere with medical
electrical devices.
The RADARMED should only be operated with the original mains cable specified in the list of contents
delivered. Operating the device with any other mains cable can lead to increased emissions or
reduced interference immunity of the device.
Guidance and manufacturer's declaration – electromagnetic interference
The RADARMED is intended for use in the electromagnetic environment specified below. The customer
or the user of the RADARMED should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/ flicker emissions
IEC 61000-3-3
WARNING
The device should not be used when placed immediately next to or stacked on top of other devices. If
operation is necessary when immediately next to or stacked on top of other devices, the device
should be monitored to ensure it is operating as intended in this arrangement.
EN109-1435752-42
Compliance
Group 1
The RADARMED uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Class B
The RADARMED is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low voltage
Class A
power supply network that supplies buildings used for
domestic purposes.
Complies
Electromagnetic environment – guidance
ENGLISH