3.3

Scan a cartridge

Use only cartidges from Miltenyi Biotec. Cartridges are for single-use only. For proper disposal of cartridges,
follow local regulations and carefully read the chapter
Do not touch the chip on the cartridge.
The cartridges are equipped with a 2D code. Scan the cartridge prior to use. Refer to the respective datasheet for
further information.
1 Locate the 2D code reader on the front of the instrument. The LED beam points downwards.
2 Hold the cartridge in front of the reader, tilting it slightly.
3 A pop-up window will appear if scanning was successful.
3.4

Load a cartridge

3
1 Open the instrument door. ( Figure 3.1 )
2 Orient the cartridge so that the Tyto owl image is in the upright position. Place the cartridge in the right slot
of the instrument stage (
rear side (
locks into position.
3 The door closes automatically. The handle flashes red, and an alarm sounds 10 seconds before closure. Hold
the door open if you need more time to insert the cartridge.
If the cartridge is installed correctly, the status bar indicator turns blue, showing
to start measurement (refer to the MACSQuantify™ Tyto® Software User manual).
36
Figure 3.1 ). To insert the cartridge, first lower the front side ( 1. ) and then lower the
2. ). Do NOT push the cartridge downward. You will hear a click as the cartridge automatically
Important safety information .
Cartridge locked: ready