Troubleshooting

Cartridge alignment fails
This may be due to one of the following:
Loaded coordinates are set incorrectly (e.g. not scanned).
l
The cartridge was inserted into the cartridge holder incorrectly.
l
Camera error (i.e. exposure set incorrectly).
l
The chip or glas layer is defect or blurred by fingerprints.
l
Camera is out of focus.
l
Software error.
l
Stage malfunction.
l
Do one or all of the following:
Reload the same cartridge.
l
Use a new cartridge.
l
Restart the software and/or the instrument.
l
If the problem persists, contact Miltenyi Biotec Technical Support.
Continuous alignment fails
Alignment tab on the orange touchscreen) is out of focus ( Figure 5.1 , A) or the fiducial
The image (in the
mark and the chip are not aligned correctly (
identifier undetected . The incorrect alignment can be caused by one of the following:
The camera is out of focus.
l
Software error.
l
Stage malfunction.
l
Do one or all of the following:
Use a new cartridge.
l
Restart the instrument.
l
Figure 5.1 , B). Sometimes, the software will give the error Tile