Technical data and specifications
Important safety information before installation and use. The instrument is not specified
Read the chapter
for use in a cold room.
The MACSQuant® Tyto® Cell Sorter is labeled as a protection class I device and must be plugged into a grounded
power outlet.
Note that for USA and Canada, the main power supply cord and plug of the instrument shall comply with the
following specifications: UL listed and KAM cord, minimum type SJ, minimum 18 AWG, 3 conductors. Rated for
a minimum temperature of 60 °C. Provided with grounding-type (NEMA 5-15P) attachment plug, rated 125
VAC, 10 A. Opposite end terminates in IEC 320-style connector, rated 125 VAC, 10A.
Supply voltage fluctuations up to +/– 10% of the nominal voltage. Transient over-voltages present on the mains
supply: category II. The instrument is suitable for rated pollution degree 2.
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy
and, if not installed and used in accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
Reorient or relocate the receiving antenna.
l
Increase the distance between the equipment and the receiver.
l
Connect the equipment to an outlet on a circuit different from that to which the receiver is connected.
l
Consult the dealer or an experienced radio/TV technician for help.
l
Changes or modifications of the instrument, unless expressly approved by Miltenyi Biotec, may void your
authority to operate the instrument pursuant to FCC 47 CFR.
For other safety considerations, refer to the product label, or visit
Design and specifications are subject to change without notice.
www.miltenyibiotec.com .