Fluidics
Sample volume
100 µL–10 mL
Purity
Up to 99%
Cell flow rate
Up to 55,000 cells/s (results depend on input sample and Poisson statistics)
Sorting rate
Up to 30,000 valve actuations/s (results depend on input sample and Poisson
statistics)
Viability
Up to 99% for lymphocytes
Sorting mode
Two-way sorting: positive cells (sorted cells) and negative cells (non-sorted cells)
Temperature control
4-25 °C (under standard environmental conditions)
Technical support
Real-time live support functionality
Pressure
Within cartridge: <210 mbar
Within system: max. 1.55 bar
Table 6.3: Fluidic parameters of the MACSQuant Tyto Cell Sorter.
Data management
Measurement parameters
Signal processing
Table 6.4: Data management.
Technical data
Dimensions
Weight
Resolution
Power requirements
Power consumption
VESA wall mount compatibility
Connections
Certification
Table 6.5: Required technical specifications and certification for external monitors for use with the MACSQuant Tyto Cell Sorter.
Height measurement for all parameters
16 bit analog-to-digital conversion and signal processing
Specification
Screen diagonal 31.5''–32''
Max. depth 80 mm
< 10 kg
1920 × 1080 at 60 Hz (optimal resolution)
Optional: 1920 × 1200
110–230 V~, 50/60 Hz
Max. 200 W
MIS-D; MIS-E
DisplayPort, HDMI
CE, UL, FCC Class B
6
49