b.
Install the J ends of the H-bracket
on the bowtie so that the H-
bracket fits into the notches on the
sides of the bowtie.
c.
Insert the cross pin into the H-
bracket, through the bowtie, and
through the other end of the H-
bracket.
WARNING:
Do not use the H-bracket without the cross pin installed correctly. To
do so could cause patient fall, injury, or equipment damage.
d.
Make sure the drop lock on the end of
the cross pin moves to the side to hold
the H-bracket in position.
e.
Repeat these steps to install an H-bracket on the other column.