Troubleshooting

Clean difficult spots

To remove difficult spots or stains, we recommend that you use standard
household cleansers and a soft bristle brush. To loosen heavy, dried-on soil,
you may first need to saturate the spot.

Steam clean

Do not use any steam cleaning device on the unit or put any Allen®
Advance Table components in an autoclave or steam cleaning system.
Excessive moisture can damage mechanisms in this unit and its
components.

Disinfect

CAUTION:
Failure to use cleaning and disinfectant products in accordance with the
manufacturer's instructions could cause equipment damage.
When there is visible soil and also between patient use, we recommend that
you disinfect the unit with an EPA registered (US only), tuberculocidal,
disinfectant such as Morning Mist Neutral Disinfectant.
Dilute the disinfectant as shown on the manufacturer's label.
Troubleshooting
WARNING:
Only Allen Medical-authorized persons should service the Allen®
Advance Table. Service by unauthorized persons could cause patient
fall, injury, or equipment damage.
If the troubleshooting information below does not fix the problem, contact
Allen Medical Technical Support "Service calls" on page 47.
The Allen® Advance Table will not go into pendant
control mode from the manual flip mode
Make sure that:
The brakes are locked.
The columns are at the same height.
The Allen® Advance Table and the installed support top are level.
See "Level the table and the support top" on page 35.
The Allen® Advance Table is free of objects that could interfere
with the adjustment.
The flip handle is in the up and locked position.
46
Allen® Advance Table User Guide (D-720722 Ver D)