Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The A-71100/A-71101 are intended for use in the electromagnetic environment specified
below. The customer or the user of the A-71100/A-71101 should make sure it is used in such
an environment.
Immunity
IEC 60601
Test
Test Level
Conducted RF
3 Vrms
IEC 61000-4-
150 kHz to
6
80 MHz
3 V/m
Radiated RF
80 MHz to
IEC 61000-4-
2.5 GHz
3
Note 1: At 80 MHZ and 800 MHz, the higher the frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the A-71100/A-71101 is used exceeds the applicable RF
compliance level above, the equipment should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the equipment.
b. Over the frequency range of 150 kHz to 80 MHz, field strength should be less than 3
V/m.
Allen® Advance Table User Guide (D-720722 Ver D)
Compliance
Electromagnetic Environment—
Level
Guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the A-
71100/A-71101, including cables, than
the recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
3 Vrms
3 V/m
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
less than the compliance level in each
frequency range
Interference may occur in the vicinity
of equipment marked with the
following symbol.
Specifications
150 kHz to
= , √
80 MHz
80 MHz to
= , √
800 MHz
800 MHz to
= , √
2.5 GHz
b
.
a
, should be
59