Clean and disinfect
We recommend that you clean and disinfect the Allen® Advance Table and
the support tops between patient use. Refer to your facility's cleaning and
disinfection policies, and follow the guidelines below.
WARNING:
The potential for electrical shock exists with electrical equipment.
Failure to follow facility protocols may cause death or serious injury.
WARNING:
The potential for electrical shock exists with electrical equipment.
Failure to follow facility protocols may cause death or serious injury.
WARNING:
Always wear the applicable personal protective equipment when you
clean the unit. Otherwise, injury could occur.
WARNING:
Do not put the unit in an equipment wash room. Patient fall, injury, or
equipment damage could occur.
WARNING:
Do not expose the unit or its components to excessive moisture. Patient
fall, injury, or equipment damage could occur.
CAUTION:
Do not use harsh cleaners, solvents, or detergents. Equipment damage
could occur.
NOTE:
Do not use iodophors; they will stain the product.
CAUTION:
If you use a quaternary-type disinfectant, make sure to rinse it off the
product immediately. Otherwise, the disinfectant could cause corrosion
on painted and plated components.
Clean
CAUTION:
Failure to use cleaning and disinfectant products in accordance with the
manufacturer's instructions could cause equipment damage.
We recommend that you clean the unit with detergent and warm water. Do
not use excessive liquid or harsh cleansers.
Allen® Advance Table User Guide (D-720722 Ver D)
Clean and disinfect