Operation (cont'd)
Read DC Voltage Outputs
Select the Read mode by pressing the SOURCE/READ button until "READ" appears on the
display. Set the toggle swtich to the READ position.
Display: V, mV
1.
Turn the Selector knob to VDC.
2.
Connect the black and red leads to the meter as shown for the appropriate range.
3.
Press the RANGE/TYPE button to select range, starting with the highest measurement
range.
4.
Connect the red (+) lead and the black (-) lead of the meter across the voltage to be
measured.
5.
Use the toggle switch to view the maximum and minimum values.
Voltage Output Signal
Controller
Transmitter
Power Supply
Connection for millivolts and volts below 10.25 VDC
Voltage Output Signal
Power Supply
Loop Voltage
Connection for volts to 200.0 VDC
26
All manuals and user guides at all-guides.com
1
4
5V MAX
45V MAX
!
!
INPUT
INPUT
!
45V
MAX
INPUT
2
3
1
4
5V MAX
45V MAX
!
!
INPUT
INPUT
!
45V
MAX
INPUT
2
3
5
!
300V
MAX
INPUT
T/C
+
5
!
300V
MAX
INPUT
T/C
+