Operation (cont'd)
Count Frequencies
Select the Read mode by pressing the SOURCE/READ button until "READ" appears on the
display. Set the toggle switch to the READ position.
Display: kHz, Hz, CPM
This function is used as a frequency counter. Available ranges are from 0.01 kHz to 10 kHz, 1 Hz
to 1000 Hz, and from 1 CPM to 1000 CPM.
1.
Turn the Selector knob to FREQ.
2.
Connect the black and red leads to the meter as shown.
3.
To measure waveforms with amplitudes between 1 V and 10.25 V rms, use the low level
inputs.
4.
Use the high voltage connection to read waveforms with amplitudes from 10.25 V to 250 V
rms.
Frequency Output Signal
Flow Meter
Flow Sensor
Variable Speed Drive
Controller
Transmitter
Frequency Output Signal
Flow Meter
Line Frequency
Variable Speed Drives