Accuracy
The Osstell Beacon instrument has an ISQ accuracy/resolution of
+/- 1 ISQ. When the SmartPeg is attached to an implant, the ISQ value
can vary up to 2 ISQ depending on SmartPeg attachment torque.
Power, weight & size:
Lithium-ion battery:
Charging:
Instrument size:
Package size:
Instrument weight:
Gross weight:
Environmental conditions during transport:
Temperature:
Relative humidity:
Pressure:
Environmental conditions during use:
Temperature:
Relative humidity:
Pressure:
IP class:
17) Troubleshooting
No measurement or unexpected value
Re-Used SmartPeg.
The SmartPegs™ are disposable and should only be used for one
or multiple measurement at one treatment session, for a single
patient only. Repeated re-use may result in false readings due to
wear and tear of the soft aluminium SmartPeg threads.
Wrong SmartPeg type selected for the Implant.
See SmartPeg reference list, osstell.com/smartpegguide
Bone or soft tissue in between SmartPeg and Implant.
Make sure to clean the Implant prosthetic connection before
attaching the SmartPeg.
Electromagnetic interference.
Remove the source of electromagnetic interference.
Instrument tip is held too far away from the SmartPeg.
Hold the instrument tip 3-5mm away from the SmartPeg.
Instrument does not sense the SmartPeg, hence no measure-
ment.
Perform a 2 cm circular movement of the instrument and then
try to measure again.
10
3.7V DC
USB cable (USB type C / USB type A)
connected to standard USB 2.0 or 3.0
(type A) outlet
210 x 35 x 25 mm
272 x 140,2 x 74,60 mm
0,07 kg
0,75 kg
-40º C to +70º C
10% to 95%
500 hPa to 1060 hPa
+10º C to +35º C
30% to 75%
700 hPa to 1060 hPa
IP20
25087-00 ML