Qualification Of User; Performance Characteristics; Maximum Electrical Capacity; Safety Instructions - Erbe 20132-046 Manual Del Usuario

Cable de conexión apc
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 62
OBJ_DOKU-189884-007.fm Seite 20 Freitag, 11. Juni 2021 10:59 10
1.4

Qualification of user

For the intended use, the product may only be operated by med-
ical professionals who have been trained in its use on the basis
of the notes on use.
1.5

Performance characteristics

Performance characteristics relating to the intended use are:
• Transporting HF current from the electrosurgical unit to the
instrument.
• Transporting the argon gas from the APC unit to the instru-
ment.
2

Maximum electrical capacity

The maximum electrical capacity of this product is:
• 4300 V
p
3

Safety instructions

WARNING!
Check insulation for damage before every application.
Erbe recommends carrying out an electrical continuity test be-
fore each application of these products.
If the cable conductor is broken or the electrical transmission of
the cable is interrupted in any other way, this may lead to arc
formation in the patient return line or in the active circuit; this
may cause burns to the patient/surgical staff or start a fire.
For an APC connector hose cable, verify continuity prior to each
application by confirming instrument recognition on the display
of an APC after connecting the cable to an APC probe and unit.