OBJ_DOKU-189884-007.fm Seite 35 Freitag, 11. Juni 2021 10:59 10
formation in the patient return line or in the active circuit; this
may cause burns to the patient/surgical staff or start a fire.
For an APC connector hose cable, verify continuity prior to each
application by confirming instrument recognition on the display
of an APC after connecting the cable to an APC probe and unit.
If damaged, do not use this product!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Damp inside the connectors may cause flash-overs that could
damage the connector or its insulation. Ensure that the product
is dried sufficiently during processing and prevent fluids from
penetrating the connector during use. The older the material, the
greater the risk of unwanted damp in the connectors.
Erbe Elektromedizin expressly warns against modifying the ca-
ble, e.g. independently replacing defective insulation. Any mod-
ification leads to the termination of liability on behalf of Erbe
Elektromedizin.
4
How to Use
Clean/disinfect and sterilize this product before using it for the
first time and before every subsequent use.
WARNING! Do not kink the cable.
WARNING! Protect this product from any form of mechanical
damage! Do not throw! Do not use force!
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the cable.
Connect the instrument to the cable. Then connect the connect-
ing cable to the unit's APC socket as follows:
• Insert the HF connector into the multifunctional socket. The
connector must click into place audibly.
35