Precleaning; Manual Cleaning And Disinfection - Erbe 20132-046 Manual Del Usuario

Cable de conexión apc
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 62
OBJ_DOKU-189884-007.fm Seite 24 Freitag, 11. Juni 2021 10:59 10
Processing step
Items Needed
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
50 ml disposable syringe with male
Luer-Lock connection (4 ea.)
Female-to-female luer-lock adapter
(=irrigation adapter)
Compressed air gun (2 bar maximum)
Machine cleaning/
Female-to-female luer-lock adapter
disinfection
(=irrigation adapter)
5.5

Precleaning

Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Screw the irrigation adapter onto the argon gas connection
of the cable.
3. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
4. Remove the syringe. Leave the irrigation adapter on the
product for the next processing steps.
5.6

Manual cleaning and disinfection

Use a liquid detergent suitable for the preparation of a dipping
bath.
Use a disinfectant compatible with the detergent suitable for the
preparation of a dipping bath.