OBJ_DOKU-189884-007.fm Seite 34 Freitag, 11. Juni 2021 10:59 10
1.3
Environment
For the intended use, the product may only be operated in prem-
ises used for medical purposes.
1.4
Qualification of user
For the intended use, the product may only be operated by med-
ical professionals who have been trained in its use on the basis
of the notes on use.
1.5
Performance characteristics
Performance characteristics relating to the intended use are:
• Transporting HF current from the electrosurgical unit to the
instrument.
• Transporting the argon gas from the APC unit to the instru-
ment.
2
Maximum Electrical Capacity
Die maximale elektrische Belastbarkeit dieses Produkts beträgt:
• 4300 V
p
3
Safety Instructions
WARNING!
Check insulation for damage before every application.
Erbe recommends carrying out an electrical continuity test be-
fore each application of these products.
If the cable conductor is broken or the electrical transmission of
the cable is interrupted in any other way, this may lead to arc
34