5.7
Cleaning and Disinfection by Machine
The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical devices
made of plastic and metal and have a pH value between 5.5 and 11.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Overview of Validated
Procedures".
1. Roll up the product loosely and lay it in a suitable rinsing
basket. In doing so make sure the product is not touching
other instruments or parts of instruments.
2. Connect the argon gas connection of the cable to a rinsing
station using an irrigation adpater.
3. Start a tested program with the following properties:
– Sufficient cleaning effect.
– Thermal disinfection (5 to 10 minutes at 194 to 203 °F
[90 to 95 °C], A
0
– Final rinse with distilled or deionized water.
– Sufficient product drying (no visible fluid residues).
4. If contaminants are still visible on the product after the end
of the routine, repeat precleaning and machine cleaning and
disinfection.
5. Disconnect the irrigation adapter from the product upon the
completion of the washing cycle.
5.8
Check
1. Check the product for visible damage and wear:
– Damage to the product, e.g. cracks, rough surface, splin-
tering.
– Damage to the product insulation and/or cable/connec-
tor insulation, e.g. cracks and breaks.
If damaged, do not use this product!
≥ 3000).