4. Use with other ventilation equipment
The Ambu ResCue Mask may be used with Bag-Valve Mask ventilation devices.
Simply detach the valve/filter assembly and attach the ResCue Mask to the device.
5. Reuse
The mask consists of a reusable mask and a single use valve. The valve cannot be
cleaned. So it needs to be discarded after used and must be replaced by a new valve
before reusing the mask.
CAUTIONS
• Don't place the used mask in the hardcase or softcase before cleaning the mask.
• After cleaning the ResCue Mask should be stored inside the case to keep
clean until next use.
6. Cleaning
The Ambu ResCue Mask is reusable and easy to clean. Hand wash cleaning procedure:
• Prior to cleaning remove the valve containing the filter by twisting and pulling.
Discard the valve. Remove and discard the elastic rubber band. Open the white
oxygen inlet port.
• Prepare the detergent bath dissolving conventional dish-washing detergent in
hot water from the tap (40-60 °C) at 1 ml per gallon (1 ml per 3.38 liters).
• Fully immerse the mask in the detergent bath placing a weight on the device to
keep it submerged in the solution for 10 minutes. Agitate the mask while it is fully
immersed to allow complete penetration of detergent to all parts of the mask.
• Thoroughly clean the mask with a soft bristle brush until all visible dirt is
removed, paying close attention to the inside of the mask and the internal union
between the cuff and the mask.
• Using a sterile syringe flush the detergent solution though the oxygen inlet
lumen. A pipe cleaner or similar may be used in hard to reach areas.
• Repeat steps 2 to 5.
• Rinse the mask for one minute with hot water from the tap (40-60 °C) ensuring
water is flushed though the oxygen inlet port.
• Let the device air-dry at room temperature for 10 minutes.
• After cleaning the ResCue mask should be visually inspected. If any damages are
seen the ResCue Mask should be discarded.
6
Clean the internal side of the case with alcohol 95% and let dry before placing the
clean mask back in the case.
CAUTIONS
• Do not boil, steam or gas autoclave this device!
NOTE
It is the user's responsibility to evaluate the effectiveness of any deviation from
the instructions provided.
7. Technical specifications
Ambu ResCue Mask is in conformity with Council Directive 93/42/EEC concerning
Medical Devices.
Connector Dimensions:
Mask:
able to be connected to a standard 22 mm male connector
Valve:
able to be connected to a standard 15 mm male connector
Dimensions: 138 mm x 113 mm x 50 mm
Weight:
Approximately 117 g
8. Material
Mask:
Polyvinylchloride
Valve:
Styrene-Butadiene Copolymers, Polypropylene and silicone
Hard case:
Polypropylene
Soft case:
Polyurethane, Ethylene Vinyl Acetate and Micro Fabric
Bag:
Nylon
9. Temperature limits
Operating:
-18 °C (-1 °F) to 50 °C (122 °F)
Storage:
Tested at - 40 °C (-40 °F) and + 70 °C (158 °F)