Pulmonary edema
Cerebral edema
13.
The Ellik Evacuator contains natural rubber latex. When it
is used with the system, take appropriate precautions to
ensure patient safety.
14. Endoscopic procedures should be performed only by persons
having adequate training and familiarity with endoscopic
techniques. Consult medical literature relative to techniques,
complications and hazards prior to performance of any
endoscopic procedure.
15. Failure to use protective filters or suitable filtering spectacles
during the activation of a surgical laser beam may cause eye-
damage to the user.
16. Temperatures near, and in front of the high intensity light
emitting tip may cause permanent tissue damage and /or
coagulation if brought into patient contact.
17. Consult literature for the safe use of electrosurgical
equipment. Inadvertent burns can occur when the appropriate
patient return path is obstructed.
18. No modification of this equipment is allowed.
19. Keep active electrosurgical probes isolated from the patient
when not in use.
20. Do not exceed the rated accessory voltage for the
resectoscope working elements:
Cut Upp = 3800 V
Spray Coag = 4100 V
5
�����������
1.
Urology hardware must be cleaned and sterilized prior to the
first use and after every subsequent use.
2.
Vaginal ultrasonography before hysteroscopy may identify
clinical conditions that will alter patient management.
3.
Intrauterine distension can usually be accomplished with
pressures in the range of 37 – 75 mmHg. Unless the systemic
blood pressure is excessive, it is seldom necessary to use
pressures greater than 75 − 80 mmHg.
4.
The surface temperature near the light cable inlet may exceed
41°C if the unit is operated at maximum brightness for
extended periods of time. Prior to disconnecting light guides
or light guide connectors, turn off the illumination source and
allow the endoscope to cool.
5.
Do not immerse or rinse instruments in cold water or any
other liquid to accelerate cooling. Forced cooling may
damage glass lenses and may render the endoscope
unusable.
6.
Devices must be pre-cleaned to avoid debris being cooked or
baked onto instrument surfaces.
EN- 7