5c. Thoroughly rinse and flush all parts and lumens with
warm, running, demineralized water.
6.
Visually inspect the hardware to ensure that it has been
thoroughly cleaned. Check joints, passages, and lumens for
remaining protein deposits by placing hydrogen peroxide on
these areas and observing for bubbles. In the case of
persistent protein deposits, repeat steps 2b through 4.
7.
Dry the hardware thoroughly with a soft towel or gauze
surgical sponge.
9
�������������
W W a a r r n n i i n n g g : :
Allow flammable solvents and agents from cleaning/ disinfecting
solutions to evaporate completely before using the hardware
with the GreenLight Laser.
Drying time depends on several variables, including altitude,
humidity, type of wrap, preconditioning, size of chamber, mass
of load, material of load, and placement in chamber. Users must
verify that drying time set in their autoclave yields dry surgical
equipment.
9.1
Autoclave System
1.
Clean and prepare the hardware as recommended in the
Cleaning section. Ensure all components are disassembled
and all stopcocks are in the open position.
2.
Sterilize the hardware using the following parameters.
W W a a r r n n i i n n g g : : Immediate-use (or "Flash") steam
sterilization is intended only for emergency situations.
Pre-Vacuum
Cycle
Immediate Use
("Flash") Pre-
Vacuum
Parameters
(unwrapped)
*Devices were validated wrapped in 2 layers of 1-ply
polypropylene wrap (KimGuard KC600 - 510k K082554)
3.
Allow the hardware to completely dry before use.
U U S S
Double wrap*
132 °C
(270°F)
4 minutes
exposure time
45 minutes dry
time
132°C
(270°F)
4 minutes
sterilization time
No dry time
E E U U
Double wrap*
134°C
(274°F)
3 minutes
exposure time
45 minutes dry
time
134°C
(274°F)
3 minutes
sterilization time
No dry time