The X8000 is equipped with Electronic Scope Sensing Technology (ESST), a
special safety feature that helps prevent accidental burns caused by a light cable
that is not connected to the scope. For more information, refer to the section
"Checking the ESST Feature". When operated with an ESST light cable, the X8000
senses when the scope and the light cable are separated and places the light source
in Standby mode. In Standby mode, the X8000 will reduce light output to a
minimum, preventing the light cable from generating excessive heat.
W
T
ARNING
LIGHT CABLE TIP WILL TAKE SEVERAL MINUTES TO COOL OFF
AFTER BEING PLACED IN STANDBY MODE AND THEREFORE MAY
STILL RESULT IN FIRE AND
AN INANIMATE OBJECT IF NOT USED PROPERLY
THE SCOPE OR THE LIGHT CABLE ON THE PATIENT OR ON THE
DRAPES OR OTHER FLAMMABLE MATERIAL
DEVICE IS IN STANDBY MODE
The Stryker X8000 Light Source consists of one of each of the following:
• light source console
• power cord
• xenon bulb module (spare or replacement bulb modules can be
purchased separately as P/N 220-201-000)
The features of the X8000 console are described below (see Figures 1 and 2).
4
3
2
1
Figure 1
Front panel of the X8000 console
1.
Power Button: Powers the unit on and off.
2.
Standby LED: Illuminates when the unit is operating in the Standby
mode. In Standby mode, light output is reduced to a minimum level.
6
,
HE SCOPE TIP
SCOPE LIGHT POST
5
,
LIGHT CABLE ADAPTER
/
OR BURNS TO THE PATIENT
,
.
6
9
,
,
USER
. D
O NOT PLACE
EVEN WHEN THE
7
8
AND
,
OR

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

220-200-000

Tabla de contenido