INSTRUCTION COULD RESULT IN FIRE OR BURNS TO THE PATIENT
USER OR AN INANIMATE OBJECT
INTERNAL SHUTTER OF THE CAMERA HIGHER IN ORDER TO RUN
THE LIGHT SOURCE AT A LOWER INTENSITY
W
T
ARNING
THE TIP OF THE SCOPE MAY EXCEED
OPERATED AT HIGH LEVELS OF BRIGHTNESS FOR EXTENDED
PERIODS OF TIME
BURNS TO THE PATIENT
Visual Display
The X8000 provides feedback through the LCD.
• Brightness: The LCD display shows the intensity level of the light as a
percentage between 0 and 100. For example, if the LCD shows "70, "
the light output to the fiberoptic cable is running at 70 percent of
capacity.
• Bulb Hours: The LCD display shows how many total hours the bulb
has operated. For example, if the LCD shows "250, " the bulb has
operated 250 total hours. The bulb hours will flash after 500 hours,
and the LCD will display "Replace Bulb" after 1000 hours.
• The LCD also displays warning and error codes. The table below lists
and defines the warning and error codes displayed.
Code
Definition
E-1
All conditions are met for the bulb to
illuminate, yet it remains off.
E-2
All conditions are not met for the bulb
to illuminate, yet it remains on.
E-3
The bulb has higher than expected
current or voltage applied to it.
E-4
The bulb is kept off because the ballast
fan is not working properly.
E-5
The bulb is kept off because the bulb fan
is not working properly.
Blinking
The bulb has exceeded its
500
recommended lifetime of 500 hours.
12
HE SURFACE TEMPERATURE NEAR THE SCOPE ADAPTER AND AT
. T
. I
N ADDITION
41ºC
HE HEATED SCOPE AND ADAPTER MAY CAUSE
,
USER OR AN INANIMATE OBJECT
,
ADJUST THE
.
IF THE UNIT IS
.
Recommended
Action
Return the X8000 for
repair.
Return the X8000 for
repair.
Return the X8000 for
repair.
Return the X8000 for
repair.
Return the X8000 for
repair.
Install a new Stryker
bulb.
,