Getting Started
Installation
Check that the pump is complete, undamaged and that the voltage rating specified on the base
plate is compatible with your AC power supply. Items supplied with this ALARIS Medical Systems
syringe pump are;
IVAC
P7000 SYRINGE PUMP
®
MULTI-POSITION POLE CLAMP
DIRECTIONS FOR USE
AC POWER CABLE (AS REQUESTED)
PROTECTIVE PACKAGING
Connect the unit to the AC power supply for 24 hours to ensure that the internal battery is fully
charged.
Should the pump fail to perform correctly, replace it in its original protective packaging and contact
a qualified service engineer for investigation.
Pole Clamp
The pole clamp is supplied fitted to the rear of the unit and will provide secure fixing to standard
I.V. poles of a diameter of up to 40mm.
The pole clamp can also be fitted in a choice of 4 fixing positions allowing the unit to be mounted
to vertical and horizontal poles, equipment rails and hospital furniture in a variety of convenient
operating orientations.
The pole clamp may be adjusted for use with horizontal fittings by using the existing fixings screws
with the alternative fixing holes in the pole clamp.
The multi position pole clamp hinge will support the unit at a range of angles on an I. V. pole. To
adjust the angle the unit needs to be secured to a pole; using a hand at either end of the
instrument, change the screen viewing and syringe access angle of the instrument.
Do not mount the unit with the AC power inlet or the syringe pointing upwards.
This could affect the electrical safety in the event of a fluid spill or lead to the
infusion of air which may be in the syringe.
Replacing the Mains Fuse
If the pump continually illuminates the battery symbol and the AC power indicator light does not
illuminate when the pump is connected to the AC power supply and switched ON, suspect that,
either the power supply fuse in the AC power plug, or, the internal fuse has blown.
First check the power supply fuse in the AC mains plug, if the AC power indicator light does not
illuminate remove the pump from service. It is recommended that the mains fuse is only replaced
by a qualified service engineer. For further information regarding the replacement of the internal
fuses refer to the technical service manual.