Enable units is used to pre-configure the dose units allowed for use by the operator when
selecting units in the Drugs and Dosing option.
7001FAOPT73 ISS 13
Configured Options
1. Press the ON/OFF power button OFF.
2. Hold down the START button and press
the power button ON.
3. The main display will show "000". Enter the
Access Code "251" using the rate selection
arrows.
4. Enter the first digit and press NEXT
indicated on the screen.
5. When the complete code has been entered
press the OK softkey indicated on the
screen.
6. Select ENABLE UNITS using the chevron
keys and press the OK softkey indicated
on the screen.
7. Use the chevrons to toggle up and down
the list of available units. Press the
MODIFY softkey to enable/disable the
units.
8. Press the OK softkey to store the units.
9. When the set-up is complete, turn the
pump OFF and return the pump into
service,
or
configuration option.
Enable Units
select
the
next
27 / 87