7001FAOPT73 ISS 13
Configured Options
14. Select the Concentration Units using the
chevron keys, press the OK softkey to
select the concentration units.
15. Enter the Default Concentration of the drug
selected, using the chevron keys to
increase
/
decrease
concentration shown on the screen. When
the
default
concentration
selected press the OK softkey.
16. Select the Minimum Concentration of the
drug selected, using the chevron keys to
increase
/
decrease
concentration shown on the screen. When
the minimum concentration has been
selected press the OK softkey.
17. Select the Maximum Concentration of the
drug selected, using the chevron keys to
increase
/
decrease
concentration shown on the screen. When
the maximum concentration has been
selected press the OK softkey.
18. Select the Bolus Rate required, using the
chevron keys to increase / decrease the
bolus rate.
When the rate has been
selected press the OK softkey.
19. Select the Maximum Bolus using the
chevron keys to increase / decrease the
volume.
When the volume has been
selected press the OK softkey.
20. Select the Pressure Alarm level using the
chevron keys to increase / decrease the
level. When the level has been selected
press the OK softkey.
21. Review the drug setup data, press the OK
softkey to confirm the data displayed. This
will return the pump to the drug setup
menu. Use the BACK softkey at any time
to return to the previous stage.
Drug Set-up
the
default
has
been
the
minimum
the
maximum
21 / 87