Display
BATTERY EMPTY
NEAR END OF SYRINGE
SYRINGE EMPTY
AC POWER FAIL
ERROR
BATTERY WAS LOW AT
LAST POWER OFF. HAS
PUMP BEEN RECHARGED
Alarm Procedures - Alarms without screen prompts
3 BEEPS
7001FAOPT73 ISS 13
Description
BATTERY EMPTY - Internal battery empty. To silence
the alarm switch the pump OFF and reconnect the
pump to AC power. Restart operation on AC power to
charge the internal battery.
NEAR END OF SYRINGE WARNING (NEOI) - This is a
timed event and can be set from 1 to 15 minutes. To set
/ change this value see the general options section in
this user manual.
SYRINGE EMPTY - END OF INFUSION (EOI) -The
pump has reached the end of the infusion. About 1%
(can be set in configuration) of the syringe volume will
remain in the syringe to minimise the risk of the infusion
of air bubbles into the set.
AC POWER SUPPLY DISCONNECTED WARNING AC
Power has been disconnected and the pump is
operating on battery power. Reconnect AC power or
press START to silence the alarm and continue battery
operation. The display will flash AC POWER FAIL. The
alarm will automatically cancel if the AC power is
reconnected.
INTERNAL MALFUNCTION - The alarm system has
detected an internal malfunction. Note the malfunction
code. Remove the unit from service for examination by
a qualified service engineer.
BATTERY MAY BE LOW/EMPTY - When the pump was
last turned off the battery was low. If the pump has not
been charged since then operate on AC power only.
Description
CALL BACK WARNING - Unit left switched ON for over
2 minutes without starting operation. Press STOP or any
of the control buttons to silence the alarm for a further 2
minutes. Alternatively, press and hold down STOP for 2
seconds to delay call back for 15 minutes.
Getting Started
Alarm Procedures
19 / 87