Configured Options
Drug Set-up
1. Press the ON/OFF power button OFF.
2. Hold down the START button and press
the power button ON.
3. The main display will show "000". Enter the
Access Code "251" using the chevron
keys.
4. Enter the first digit and press NEXT
indicated on the screen.
5. When the complete code has been entered
press OK softkey indicated on the screen.
6. Select the DRUG SETUP option using the
chevron keys and press the OK softkey
indicated on the screen.
7. Select the required drug with the chevron
keys and press the OK softkey.
8. To use a drug it must be enabled. Press the
YES softkey indicated on the screen to
enable the selected drug.
9. To add or change a drug name use the
chevron keys to toggle through the
alphabet. To select a letter press the NEXT
softkey. On completion press the OK
softkey indicated on the screen.
10. Select
the
Dose
maintenance doses. Use the chevron keys
to select the dose units required. When the
dose units have been selected press the
OK softkey.
11. Select the Default Doserate using the
chevron keys to increase / decrease the
doserate displayed. When the required
doserate has been selected press the OK
softkey.
12. Select the Minimum Doserate using the
chevron keys to increase / decrease the
doserate displayed. When the minimum
doserate has been selected press the OK
softkey.
13. Select the Maximum Doserate using the
chevron keys to increase / decrease the
doserate displayed. When the maximum
doserate has been selected press the OK
softkey.
20 / 87
Units
required
for